(2) Low-pressure Gradient System
Mobile phase
Detergent
Chromaster
organizer
Chromaster 5110
pump
Degassing unit
Chromaster 5440
fluorescence detector
Chromaster 5310
column oven
Chromaster 5210
autosampler
Drain
Mixer
Proportioning valve
Column
Detector cell
Four-way joint
Injection valve
Wash pump
Appended Fig. 1-9 Piping Arrangement for Low-pressure
Gradient System
Connect the inlet tube of the mobile-phase container to the
IN port of the degassing unit. Connect the OUT port of the
degassing unit to the IN port of the proportioning valve.
Connect the OUT port of the proportioning valve to the four-
way joint.
Connect the mixer of the pump to port 1 of the injection
valve of the autosampler. Connect port 2 of the injection
valve to the column inlet. Connect the column outlet to the
inlet tube of the detector cell. For detergent, provide
connection to the three way solenoid valve of the syringe
valve of the autosampler through the degassing unit.
APPENDIX 1 - 13