2.2
(c) At occurrence of pressure sensor error
* Active status (error detected) can be checked from
the Chromato Data Station (CDS) or GUI controller.
(8) Material of Wetted Parts
TEFLON-AF (degassing membrane), glass fiber-reinforced
PPS (connector)
(9) Power
Supply
24 V DC
(10) Power Consumption
Max. 12 W (normally 9 W)
(11) Dimensions
53 (W)
×
297 (D)
×
100 (H) mm
(12) Operating Principle
The liquid to be degassed is flowed through a special
synthetic resin membrane placed in the vacuum chamber of
6-channel degassing unit. The molecules of gases
dissolved in liquid are small, and have a higher mobility
than the liquid and a high affinity to the resin membrane.
Therefore, the gases are selectively transmitted through the
membrane and discharged out of the degassing unit.
Thus, the gases are eliminated from the liquid.
2 - 3