Table 3-2
Reagents for Mobile Phase that may be Restricted
Usable reagents
Usable, if it is 50%
or lower
Usable, if it is 10% or
lower
Unusable reagents for
SUS316
Phosphoric acid
Sodium phosphate
Acetic acid
Ammonium citrate
Ammonium nitrate
Citric acid
Butyric acid
Sodium nitrate
Disodium phosphate
Ammonium formate
Ammonium perchlorate
Ammonium phosphate
Boric acid
Formic acid (up to pH3)
Hydrochloric acid (up to
pH3)
Potassium nitrate
Sodium
hydrogencarbonate
Sodium carbonate
Ammonium chloride
Potassium chloride
Sodium chloride
Tolufluoroacetic acid
Note 1
:
Max. concentration of
above reagents
should be 5-6%.
Note 2
:
When the above
reagents have been
used necessarily,
wash all the flow
paths with distilled
water completely
without fail.
Do not use reagents that corrode SUS316 though they are not
shown in the table above. High viscosity solvent won’t be
delivered, so please be careful.
Table 3-3 Isocratic Analysis
One Fluid Wash <standard>
(for Cut injection method,
Full loop injection method)
Two Fluids Wash
(for All volume injection method,
Low carry over analysis)
Wash fluid 1
Sample solution
Mobile phase
Sample solution
(Strong wash)
Wash fluid 2
Aqueous solvent
Mobile phase
(Weak wash)
Table 3-3 Gradient Analysis
One Fluid Wash <standard>
(for Cut injection method,
Full loop injection method)
Two Fluid Wash
(for All volume injection method,
Low carry over analysis)
Wash fluid 1
Sample solution
Sample solution
Gradient final solvent
(Strong wash)
Wash fluid 2
Aqueous solvent
Gradient initial solvent
(Weak wash)
3 - 63