3.1 Turning on Power Supply
3.1.3 Auto
Zero
A/Z
(Auto zero)
ABS indication and output signal on the monitor screen can be
zeroed.
0.0 250 0.0000
1
D2
TIME
WL
ABS
PROG
LAMP [I]
Auto zero adjustable range: -0.2 to 2 AU
If this range is exceeded, the following message appears.
PRESS “CL” KEY TO CLEAR MESSAGE
A/Z OVER RANGE
NOTE
: Auto zero function is unusable during 2-
wavelength measurement.
3.1.4 Setting of Measuring Wavelength
A measuring wavelength is settable.
WL
(Wavelength)
The measuring wavelength setting screen appears.
250
D2 MODE (190 – 600), W MODE (380 – 900)
WAVELENGTH (190 – 900) nm
Press [numerical] keys (within [1][9][0] to [9][0][0]), and then the
[ENT] key.
* Settable wavelength range: 190 to 900 nm, enterable in
1-nm steps
3 - 8