3.4.1
3.4 Analysis Operation
3.4.1 Time Program Setting
In the Chromaster 5310, a time program can be prepared for
each of the temperature setting and selector valve position
setting.
SET PROGRAM NO. (1 - 9)
1
EDIT MODE (MODIFY=0
,
NEW=1)
0
TIME TEMP VALVE1
0.0 1
TIME TEMP VALVE1
0.0 _
1
TIME TEMP VALVE1
0.0 40
_
1
TIME TEMP VALVE1 EVENT
_ 1
Enter a time program number.
1 ENT
Select a mode.
•
MODIFY : For editing an existing time program.
•
NEW
: For deleting an existing time program and
creating a new one.
0 ENT
Indicates the number
assigned to the time program
being edited.
Time setting
Enter a preset time.
ENT
Temperature
setting
Enter a preset temperature.
40 ENT
Valve
setting
2 ENT
PROG
New step
For the 2nd step onward, it is not allowed to enter 0.0 min.
On completion of program settings, press the ESC key.