Symptom Cause
Judgment/Check
Remedy
The warm-up period is
not sufficient.
⎯
Wait until the instrument
becomes stable (at least
10 minutes).
Mobile phase is leaking
from a fitting.
Check each fitting.
Retighten the offending
fitting.
An impurity is eluting
from the column.
Stop feeding liquid
from the pump.
Wait until elution is
completed, or replace the
column with a new one.
Excessive drift
There is a significant
change in the ambient
temperature. Or low
purity solvent is used.
Check solvent.
Control the temperature
variation.
Purify the solvent or
replace it with one of a
higher quality.
Flow cell contaminated
Visual check
Clean the flow cell
(see 4.4).
Xenon lamp off-position
Check lamp energy
level
Adjust lamp position
(5.2.1).
Error in flow cell
mounting
Visual check
Correctly mount flow cell
(4.4).
Poor S/N ratio
Flow cell surface
scratched
Visual check
Replace flow cell (4.4).
Water is not injected in
the flow cell.
⎯
Inject water into the flow
cell.
Bubbles have entered
the flow cell.
Visual check
Remove bubbles (4.4).
Poor adjustment of
xenon lamp position
Check lamp energy
level.
Adjust Xe lamp position
properly.
Error in flow cell
mounting
Visual check
Correctly mount the flow
cell (4.4).
Poor wavelength
accuracy
Monochromator off-
position
Check wavelength
accuracy.
Perform wavelength
accuracy (4.3).
Liquid has leaked.
Visual check
Wipe off the liquid which
has leaked to the drain
receptacle. Also wipe off
the liquid stuck to the leak
sensor (4.2.5).
Leak error occurs
Contaminant has stuck
to the leak sensor.
Visual check
Wipe off the contaminant
stuck to the leak sensor
(4.2.5).
4 - 32