(3) Put a new injection port seal into the injection port.
Attach it to the injection valve, and tighten it up with your
hand.
Turn the taper part of the injection
port seal to the injection side when
attach it.
Injection port seal
Taper part
Injection port
Fig. 5-3
Attachment of Injection Port Seal
NOTES
: 1. Turn the taper part of the injection port seal to
the injection port side when attach it.
2.
Tighten the injection port with your hand until
it is finger-tight.
3.
Avoid using tools to prevent the injection port
seal from deformation.
(4) After replacing the injection port seal, press the
WASH
key
to carry out injection port cleaning.
NOTES
: 1. After replacement of the injection port seal U,
make sure liquid is not leaking through the
seal.
2. If the tip of needle is deformed, the service
life of port seal may be significantly
shortened. Use the needle free from tip
deformation.
5 - 5