(3) Sampling
The syringe aspirates one-half of the air volume (:va/2) while
needle is moving to the sample vial.
The syringe aspirates [vw + vi :Waste Injection
volume] after the needle has moved into the sample vial.
Buffer tube
vd :Dead volume (the instrument inside parameter)
vi :Injection volume (the sample net volume to the flow path)
vw :Waste volume (the air volume of the lead and rear part of a sample)
va :Air volume (the air volume which sample is sandwiched.)
Injection valve
Syringe valve
Syringe
Wash pump
5
1
3
2
4
Plunger
wash
vw + vi
Injection port
5
4
3
2
1
6
Waste
solution
Loop
Waste
solution
Wash port
Sample
Rack
P
Needle
va/2
va/2
From Pump To Column
Wash
fluid2
Wash
fluid1
Degassing
unit
3-way solenoid valve
2 - 38