(2) Sequential program setting
(a) Setting
method
As a method of setting a sequential program, the
setting procedure is shown below on the following
conditions; analysis time 5 minutes and injection of
120 samples of 10
μ
L each sequentially once.
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0 [C]
This display appears at the end of initialization after power-
on.
PROG
EDIT MODE (MODIFY=0, NEW=1)
0
Select MODIFY for editing the previous program or NEW for
creating a new program.
0 ENT or 1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 1 5.0
Input a rack code.
1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 1 5.0
Input the first vial number.
1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 1 5.0
Input the last vial number.
120 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1-200 10.0 1 5.0
Input an injection volume (
μ
L).
10 ENT
(to be continued)
3 - 13