Table 4-3 Autosampler Main Unit Errors (continued)
No.
Error Message
Possible Cause
Remedy
13
INJ TIMING ERROR
1. Any pump other than the
Chromaster 5110 pump is
combined for use with the
autosampler.
2. The flow rate specified for
the pump is too low.
1. Set the Injection Timing
function to “NO USE”.
2. Adjust the minimum flow rate
properly. Alternatively, set
the Injection Timing function to
“NO USE”.
1. No sample vials are set.
2. The clearance between
the top face of sample vial
and the detection lever is
inappropriate. Or the
detector lever is caught.
1. Set sample vials.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
14 NO
TUBE
(Note: In execution of
a sequential program,
it is terminated
without indication of
“NO TUBE” on the UI
pad. In execution of
the TUBE command,
“NO TUBE” is
indicated on the UI
pad. Refer to
Section 3.3.2.)
3. The electric system
malfunctions.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
15 STOPPED
BY
EXTERNAL ERROR
An error has occurred in other
unit connected.
Eliminate the cause of the error to
release the error status.
16 SEQ
PROGRAM
BUFFER
OVERFLOW
Communication is abnormal.
Contact your nearest service
office of Hitachi High-
Technologies Corporation sales
representative.
1. The sample rack or
sample vial is blocking
movement.
1. Set the sample rack or sample
vial correctly.
2. There is an obstacle on
the moving route of arm X
(in left-right direction).
2. Remove the obstacle.
17 ERROR-X
3. The electric system
malfunctions.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. There is an obstacle on
the moving route of arm Y
(in front-rear direction).
1. Remove the obstacle.
18 ERROR-Y
2. The electric system
malfunctions.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. There is an obstacle on
the moving route of needle
Z (vertically).
1. Remove the obstacle.
19 ERROR-Z
2. The electric system
malfunctions.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.