(c) HEIGHT
The needle can be moved to the specified height.
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0
FUNC
FUNC (TUBE=1, MOVE=2, HEIGHT=3, SYR=4,
3 VALVE=5, SVALVE=6, SPURGE=7, WPUMP=8)
3 ENT
ESC
HEIGHT ( Z SPEED)
<0.0-53.0> 0.0 2
HEIGHT ( Z SPEED)
<1-2> 14.5 2
e.g.) 14.5 ENT
MOVING [B]
WAIT!
2 ENT
IDLE
PRESS FUNC KEY OR IDLE KEY [W]
FUNC
ESC
ESC
Input a desired height.
Input a needle moving speed.
1: The needle will move at the low speed. Select this speed
for using a hard septum, etc.
2: The needle will move at the high speed. Usually, this
speed is used. However, for moving the needle upward,
the needle will move at the high speed irrespective of the
setting made here.
The needle moves to the specified height.
MOVING [B]
IDLE WAIT!
Monitor screen reappears.
3 - 44