5.3 Replacement of Injection Port Seal
The injection port seal should be replaced periodically every
about 10,000 sample injections as a standard.
Items to be prepared:
Injection port seal (P/N 893-0810)
Seal remover (P/N 893-2115)
(1) Loosen the injection port with your hand. And remove it
from the injection valve.
Injection port
Injection valve
Fig. 5-1 Removal of Injection Port
(2) Insert a seal remover into the injection port. And remove
an injection port seal.
Injection port seal
Injection port
Insert a seal remover into
the injection port.
Remove an injection port seal.
Seal remover
Fig. 5-2
Removal of Injection Port Seal
5 - 4