5. PARTS REPLACEMENT
5.1 Consumables and Spare Parts
5.1.1 Consumables
The following parts are consumables. They must always be on
hand. The amount of consumption of each item is based on the
assumption that the autosampler is operated 240 days per year,
8 hours per day, and 15 minutes per analysis. “Standard life”
indicates the number of uses under running of pure water. This
value varies widely with operating conditions. Therefore, each
item should be prepared with a margin in quantity depending on
the actual operating conditions.
NOTE
: Please replace service parts and consumables in this
instruction manual by the specified cycle. It may
causes malfunction of the instrument without replacing
them.
Table 5-1 Consumables
Part No.
Name
Form / Specification
Location
(Estimated Life)
Standard
Life
080-3595
Sample vial
(500 pcs)
Sample vial
⎯
855-2351 Septum
(100 pcs)
slit(
-
)
Sample vial
⎯
080-3704 Cap
(500 pcs)
Sample vial
⎯
810-1337 Solvent
filter
Wash fluid bottle One year
893-0810 Injection
port
seal
Injection port
(One year)
10,000
injections
893-0811 Injection
valve stator
Injection valve
(Two years)
20.000
switchings
32
φ
11.5
φ
2.5
φ
34
5 - 1