3 - 14
SET PROGRAM: No. ( 1 – 9 )
1
EDIT MODE (MODIFY = 0, NEW = 1)
0
Select a program number from 1 to 9.
ENT
PRESSURE (MAX MIN)
(0.0 – 40.0) 20.0 0.0 1
Select EDIT MODE
ENT
Press the key to change the mode from manual set
screen to the programming mode.
Program generation is started
1
INITIAL
PROG
TIME FLOW EVENT
0.0 1.000 12 22 32 42 1
TIME FLOW EVENT
10.0 0.500 1
Input TIME, FLOW rate, and EVENT.
Up to 4 entries can be made per step for event.
Refer to 3.2.8 for details of EVENT.
ENT
TIME FLOW EVENT
1
A new input screen will be appeared.
Input TIME, FLOW rate, and EVENT.
ENT
Enter MAX and MIN pressure limit.
ENT
A new input screen will be appeared. Press the ESC key to
close the input screen, and return to the monitor screen.
ESC
Monitor screen