(7) Sample
loading
Deliver the sample into the sample loop.
The syringe discharges [va/2 + vw/2 + vi + vd : one-half of the
Air one-half of the Waste Injection volume
+ Dead volume (inside parameter)], and delivers a sample
into the loop.
Buffer tube
va/2 + vw/2
+ vi + vd
vd :Dead volume (the instrument inside parameter)
vi :Injection volume (the sample net volume to the flow path)
vw :Waste volume (the air volume of the lead and rear part of a sample)
va :Air volume (the air volume which sample is sandwiched.)
Syringe vlave
Syringe
Wash pump
5
1
3
2
4
Plunger
wash
Sample
Rack
P
Needle
Injection valve
Injection port
5
4
3
2
1
6
Loop
vw/2
vw/2
Wash
fluid2
Wash
fluid1
Waste
solution
Waste
solution
Wash port
From Pump To Column
Degassing
unit
3-way solenoid valve
2 - 42