3 - 61
3.5 Setup
3.5.1 Pump Mode Setting
This section shows how to set the pump in the single-liquid
delivery mode (FLOW mode).
NOTE
: Once this setting is made, it is memorized and need not
be made each time.
(to be continued)
SETUP: PUMP MODE CONTRAST S.NO
1 FLOW 4 02AB-111
SETUP ITEMS (PUMP=1 CONTRAST=2 D.MIX=3)
1
ENT
PUMP (1=1, 2=2, 3=3)
1
Select PUMP
1 ENT
SET UP
UTLTY
POWER ON +
Select PUMP number
3 ENT
GR MODE (LOW=1, FLOW=2)
2
Select FLOW (isocratic mode)
2 ENT
LOW :Low gradient mode
FLOW :Isocratic mode
Press the POWER and SET UP UTLTY keys simultaneously.
The Pump mode setting screen will be displayed.
Continue pressing SET UP UTLTY until the setting screen appears.
SETUP: PUMP MODE CONTRAST S.NO
3 FLOW 4 02AB-111
1 ENT
NOTE
: Only FLOW (isocratic mode) is selectable for the
PUMP3 setting.