4.6
4.6 Error
Messages
Message Description
Remedy
ROM ERROR
PRESS “CL” KEY TO CLEAR
MESSAGE
ROM is faulty.
Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
RAM ERROR
PRESS “CL” KEY TO CLEAR
MESSAGE
RAM is faulty.
Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
PARAMETER ERROR
PARAMETERS ARE SET TO
DEFAULT VALUE.
Parameter is wrong.
1) Turn on power supply again.
2) Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
LOG INFORMATION ERROR
LOG INFORMATION IS
INITIALIZED.
Log information is wrong.
1) Turn on power supply again.
2) Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
EEPROM ERROR
INSTRUMENT INFO. IS
INITIALIZED.
Instrument information is
wrong.
1) Turn on power supply again.
2) Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
WL DRIVE MECHANISM
ERROR
Initialize test is
unsuccessful.
1) Clear the message by CL key.
2) Confirm proper mounting of the
flow cell, and then turn on power
supply again.
3) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
WL MOTOR DRIVE DRIVER
ERROR
Motor driver for
wavelength is faulty.
1) Turn on power supply again.
2) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
WL CALIBRATION ERROR
PRESS “CL” KEY TO CLEAR
MESSAGE
The 656 nm bright line of
D
2
lamp is not found in
self-diagnosis (at
initialization).
1) Clear the message by CL key.
2) Confirm proper mounting of the
flow cell, and then turn on power
supply again.
3) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
THERMO CELL ERROR
Thermostatic flow cell is
faulty.
1) Confirm proper mounting of the
thermostatic flow cell, and then
turn on power supply again.
2) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
THERMO CELL IS NOT
CONNECTED
PRESS “CL” KEY TO CLEAR
MESSAGE
Thermostatic flow cell
cannot be connected.
1) Clear the message by CL key.
2) Confirm proper mounting of the
thermostatic flow cell, and then
turn on power supply again.
3) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
4 - 21
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......