3.1 Basic Operation
3.1.4 Setting of Temperature Ready Conditions
Make the condition setting for the temperature ready status.
Initial setting: TOLERANCE=1.0 °C, WAIT TIME=0 min.
COND
NOTE
: 1. Unless a tolerance range and wait time are entered
(and ENT key clicked) on the temperature ready
condition setting screen, the settings will not be
updated.
2. If “NO” is selected for “USE TOLERANCE”, the
temperature ready status is provided regardless of
the current oven temperature.
TOLERANCE (
°
C)
WAIT TIME (min)
<0.1 - 9.9>
1.0
5
TOLERANCE (
°
C)
WAIT TIME (min)
<0 - 99>
1.0
5
Set a tolerance range.
1.0 ENT
Set a wait time.
5 ENT
Monitor screen
USE TOLERANCE (YES=1, NO=0)
0
Specify whether or not to
use tolerance setting.
1 ENT
COND