3 - 21
Pressing the purge key in the state of PUMP ON executes
purging at the set flow rate. Re-pressing the purge key during
purge will reset it. During purge, the upper pressure limit is set
at 2.9 MPa and the lower pressure limit at 0 MPa.
3.2.4 Flow Rate Calibration Setting
Set up a calibration coefficient for the flow rate.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
4 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
CALIBRATION (FLOW=1, PRESS=2)
1
Select CALIBRATION
4 ENT
SET UP
UTLTY
Monitor screen
Select FLOW
1 ENT
FLOW CALIB (0.970 – 1.030)
1.000
Monitor screen
Enter calibration coefficient.
ENT