2 - 8
No. Key
Function
Remarks
7
INSERT
Used for program editing.
Inserts a new row as last row of a program.
8
1. Press this key once to return the
program by one step.
2. Selects the Monitor display etc.
9
1. Press this key once to advance the
program by one step.
2. Selects the Monitor display etc.
10 START
Starts the gradient program.
Start is possible only when the Pump is
working, and the screen is displaying
"Monitor".
If the screen is not
displaying "Monitor", press
the ESC key to display
the "Monitor" screen, then
press this key.
11 STOP
Stops the gradient program. Stop is
possible only when the Pump is working,
and the screen is displaying "Monitor".
If the screen is not
displaying "Monitor", press
the ESC key to display
the "Monitor" screen, then
press this key.
12 MANUAL
Selects, displays, and sets Manual
operation (mixing ratio of mobile phase,
FLOW, P-MAX and P-MIN).
The conditions before
setting are held until the
parameter setting is
finished.
13 LOCK
OFF
1. Returns the screen to Monitor from
Numerical value input mode for inputting
each mode.
2. Releases key lock.
14
15
16
17
A
B
C
D
Press the "A" key in Manual status to
obtain % A = 100 in one touch. B, C or D
has a similar function.
This key can be used
depending on the gradient
conditions.
Low-pressure gradient:
A, B, C, D, key
18
Inputs a decimal point.
It is not necessary to input zero before this
point.
19
Deletes each step when editing the gradient
program.
When a cursor is in an input part of TIME:
deletes the step.
When a cursor is in an input part besides
TIME: deletes the item indicated by the
cursor.
Cancellation of the DEL key process:
Press the ESC key while “-” is indicating. It
returns to a monitor screen, and the delete
process will be cancelled.
Time program of a Pump and Detectors:
Press the DEL key. “-“ will be indicated. And
press the ENT key to determine.
Sequential program of an Autosampler:
Press the DEL key to delete it including the
step number. “-“ will be indicated. And press
the ENT key to determine.
▽
△
PROG
7
8
0
ESC
.
DEL