3.5 Setups
3.5.2 Setting of Thermostatic Flow Cell
Press and keys simultaneously.
UTLTY
SET UP
POWER
The SETUP screen now appears.
SETUP:
DETECTOR
THERMO
CONTRAST S.NO.
CH1 NO
USE
4
1401-001
Press the [ENT] key.
The selection screen appears.
SETUP ITEMS (DETECTOR=1, THERMO CELL=2,
1
CONTRAST=3)
Press numerical key [2], and then the [ENT] key.
THERMO CELL (NO USE=0, USE=1)
0
For avoiding use of
thermostatic flow cell : Press [numerical] key [0], and then the
[ENT] key.
For selecting use of
thermostatic flow cell : Press [numerical] key [1], and then the
[ENT] key.
After setting use or non-use of the thermostatic flow cell, “ON” or
“OFF” is indicated on the monitor screen 2.
WL SAMP
REF
CELL
[I]
250 301000 492000 ON
When the thermostatic flow cell is not yet connected,
“---“ will be indicated on the monitor screen 2.
WL SAMP
REF
CELL
[I]
250 301000 492000 ---
New setting is accepted after turning on power supply again.
NOTE
: Set the thermostatic flow cell in accordance with
measuring conditions. (Recommended flow rate
range when using thermostatic flow cell: 2mL/min
or less).
3 - 46