5.3 Assembling of Pre-heating Tubing
Chromaster 5310 column oven has a pre-heating section for
preliminary heating (Appended Fig. 1-8). The pre-heating
section has nine grooves (lines) for accommodating the pre-
heating tubing, and the optimum number of grooves to be used
varies depending on the flow rate for the analysis to be used.
Make optimum pre-heating, referring to Table 1 shown below.
<Preparation items>
•
Hexagon screw wrench (P/N: N315004, Size: B3)
Pre-heating
section
Appended Fig. 1-8 Front Elevation View of Column Oven
(with the Door Open)
Appended Table 1-5
Relationships between Operating Flow Rate and Optimum
Number of Lines to Be Used for Pre-heating (for SUS Tubing of
0.25 mm in Inner Dia.)
Applied Range of
Flow Rate (mL/min)
0.7 or
less
0.71 to
1.5
1.51 to
2.2
2.21 to
3.0
3.01 to
3.7
3.7 or
more
Number of pre-heating lines
0.5
1
1.5
2
2.5
3
Total length of tubing (mm)
700
800
900
1000
1000
1200