2.3 Rear Panel
2.3 Rear
Panel
PROCESSOR RECORDER MARKER IN
2 3
4
5
6
Fig. 2-3 Rear Panel
No. Indication
Function
Remarks
(1) Used to transmit the
acquisition signal (or acquired
spectral data) via the e-Line.
Also, used to send/receive the
instrument condition data via
the e-Line.
(1) Absorbance values are
output.
1 e-Line
(2) Used for input/output of each
contact signal.
(a) START IN
(b) ERROR IN
(c) ERROR OUT
(d) BUSY OUT
(2)- (a) Program start
(b) Recognition of error
in other instrument
(c) Indication of error
condition
(d)
Indication
of
operating status
2
LAMP OFF IN
The light source lamp is turned off
in response to contact signal input.
3
AUTO ZERO IN
Auto-zero operation is executed in
response to contact signal input.
4 PROCESSOR
(option)
Analog output for data processor
(integrator)
(1) Absorbance value at each time
point
(2) Stored spectral data
Absorbance value is output.
5 RECORDER
(option)
Analog output for recorder
(1) Absorbance value at each time
point
(2) Stored spectral data
Absorbance value is output.
6 MARKER
IN
(option)
The marker function is started in
response to contact signal input.
This function is started
when the short-circuit
signal lasts 1 sec or longer.