5.2.2
(5) Check the energy of W lamp. (Refer to 4.4.1.)
(6) Input the date of W lamp replacement as follows.
(a) Press the (GLP) key, and the following screen will
appear.
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
GLP (LOGBOOK=1, KEY LOCK=2,
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
1
LAMP SELECT (D2=1, W=2, Hg=3)
(c) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
1
LOGBOOK (REPORT=1, RESET=2)
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
<1–12>
_9 15 2010
W LAMP CHANGE (MM DD YYYY)
(e) Key in the date of replacement in the format of month-
day-year. For instance, if the lamp is replaced on
March 1, 2010, key in the date as shown below.
[3]
→
[ENT]
→
[1]
→
[ENT]
→
[2][0][1][0]
→
[ENT]
Thus, the date of lamp replacement is stored in
memory. At the same time, the cumulative turn-on
time of the W lamp is cleared to zero (0), and the turn-
on count is reset to 1.
NOTE
: The logbook data is reset when the numeric value
of the year and [ENT] are keyed in.
Therefore, even if the same year as that of
replacement is indicated, the year must be
numerically input followed by pressing the [ENT]
key.