Select Hg.
Press numeric key 2 and then ENT key.
(4) The following display will be presented.
<Logbook selecting screen>
LOGBOOK
(
REPORT=1, RESET=2
)
2
Select RESET.
Press numeric key 2 and then ENT key.
Next, advance to the Hg lamp change date setting screen
for setting the date of mercury lamp change.
(5) Set the date of mercury lamp change in the following
procedure.
(a) Input a month for MM (cursor appears under MM).
<Hg lamp change date setting screen (month)>
Hg LAMP CHANGE
(MM DD YYYY)
<1-12>
_9 1 2010
Press numeric key (1 to 1 2) and then the ENT key.
The cursor moves to DD.
(b) Input a day for DD.
<Hg lamp change date setting screen (day)>
Xe LAMP CHANGE
(MM DD YYYY)
<1-31>
9 1 2010
Press numeric key (1 to 3 1) and then the ENT key.
The cursor moves to YYYY.
(c) Input a year for YYYY.
<Hg lamp change date setting screen (year)>
Xe LAMP CHANGE
(MM DD YYYY)
<2000-2100>
9 1 2010
Press numeric keys (2 0 0 0 to 2 1 0 0) and then the
ENT key.
3 - 22