♦
Typically, the negative input limit for an integrator or data
processor is approximately -10 mV.
If the baseline decreases below this level, it becomes
impossible to integrate the peaks in the chromatogram
(see Fig. 3-2 (a)).
In such a case, the offset function is used to adjust the
shifted baseline to the allowable input signal range of
the data processor (see Fig. 3-2 (b)).
0
-10 mV
(a) Offset is not used
(b) Offset is used
Fig. 3-2
Example of Offsetting
(4) Offset setting is now completed and the 1
st
monitor screen
returns.
Though no value is indicated for FL, “+*” will be displayed
when selecting a value other than 0.
<1st monitor screen>
TIME Ex Em FL
PROG LAMP [I]
0.0 250
350 0.0+
∗
1
Xe
3.2.2 Time constant setting
Specify electrical response time.
(1) Press
the
UTLTY
SET UP
key.
(2) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_2
USE PROG=3, PMT VOLT=4, OTHERS=5)
3 - 9
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......