Set a photomultiplier voltage.
Press numeric key (1 to 5, usually input 3) and then the
ENT key.
* Default value: 3
(4) Photomultiplier voltage setting is now completed and the 1
st
monitor screen returns.
3.2.5 Other
settings
Set the following items.
•
Em
BANDWIDTH
•
ERROR
OUT
•
LEAK
SENSOR
•
LAMP
(1) Em BANDWIDTH setting
A bandwidth value of the emission monochromator is
settable. Standard is 15 nm and wide setting is 30 nm.
(a) Press
the
UTLTY
SET UP
key.
(b) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OTHERS.
Press numeric key 5 and then the ENT key.
(c) The following display will be presented.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
1
LEAK SENSOR=3, LAMP=4)
Select Em BANDWIDTH.
Press numeric key 1 and then the ENT key.
3 - 12
Summary of Contents for CHROMASTER 5160