5 - 11
(4) If a clogged filter has to be replaced, allow the filter has to
be replaced, allow the filter assembly to dry, then place it in
a freezer (-10
℃
) for about five minutes. The cold
temperature will contract the Teflon
®
packing and facilitate
removal of the filter.
(5) Carefully push and mount the new filter into the flow path
filter.
(6) Before attaching the flow path filter to the drain valve, clean
the flow path by running the pump.
(7) Attach the flow path filter to the lower right (canted) part of
the drain valve. Align the flow path filter with the drain valve
and tighten it securely with the wrench 21.
NOTE
: Poor tightening may cause leaks.
Flow path filter
Freezer (-10 °C about 5 minutes)
Filter
638-1423
Seal
638-1422
Flow path filter
Packing(PEEK)