APPENDIX 1 - 3
(f) Shall be free from excessive dust.
Otherwise, performance degradation might be caused.
(g) Shall be free from a sudden change of power supply
voltage. Otherwise, noise might be caused.
(h) Power supply to motor-driven equipment (such as
stirrer and vibrator) without a noise suppressor shall
not be turned ON/OFF frequently on the same power
supply line as that of this instrument.
3. Items to be Prepared by Customer
3.1 Power Supply Arrangement
(1) Power supply voltage: 100 to 240 V AC
Fluctuation shall be ±10% of the rated voltage.
(2) Frequency: 50 or 60 Hz
Fluctuation shall be ±0.5 Hz of the rated frequency.
(3) Power Capacity:
100 VA or more for a single unit and
600 VA or more for the total system
The total system shown comprises an organizer,
Chromaster 5110/5160 pump, Chromaster 5210/5260
autosampler, Chromaster 5310 column oven, Chromaster
5410/5420/5430/5450 detector.
Confirm the necessary power capacity taking actual system
configuration and OA equipment in the installation room
into account and prepare a power supply having an
adequate power capacity.
(4) Power
outlet:
Power supply shall be obtainable near the place for
installing the system inclusive of this detector.
This instrument requires one or more grounded power
outlets (3-pole type). One or more outlets are needed for
the entire system (see the section 5.1 of the Appendix.)
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......