3.4 Analyzing Operation
(b) Storing the Background Spectral Data in Memory (“2” key
followed by ENT key)
Scan is automatically carried out.
During acquisition of the background spectrum, the
following screen is presented.
Upon completion of spectral data acquisition, the monitor
screen automatically appears.
(c) Storing the Reference Energy with D
2
lamp in Memory (“3”
key followed by ENT key)
Set the minimum wavelength of scan range.
Press [numerical] keys (within [2][0][0] to [8][5][0]), and then
the [ENT] key.
* Enterable range: 200 to 850 nm, enterable in 1-nm
steps
Set the maximum wavelength of scan range.
Press [numerical] keys (within [2][5][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: Minimum scan range is 50 nm.
Measuring span is max. 400 nm.
Input the number of spectral data.
WL: *** nm
SCANNING WL RANGE (WL1 WL2) nm D2 REF
<200 – 850>
200 – 600
SCANNING WL RANGE (WL1 WL2) nm D2 REF
<250 – 600>
200 – 600
MEMORIZE SPECTRUM NO. (1–3)
_
SCANNING BACK GROUND SPECTRUM
3 - 32