3 - 37
⑤
Conditioning
time
During the conditioning time, liquid is delivered under the
initial condition of execution step (program No. 2 in this
example).
⑥
Initial
Liquid is delivered in the initial status (initial condition of
program No. 2 in this example).
⑦
Gradient
When “start” is input in the initial status, gradient elution
(program No. 2 in this example) will start.
⑧
Conditioning
time
After completion of the final step of link program, the
system returns to the first step (program No. 1 in this
example) and conditioning is carried out.
⑨
Delay time (initial)
Liquid is delivered in the initial status (initial condition of
program No. 1 in this example).
When “start” is not input during the delay time, column
flushing in
⑩
will begin.
When “start” is input during the delay time, the system
returns to action in
④
.
⑩
Column flushing
TIME %A %B %C %D
FLOW PRESS
[
LC
]
0.0 30 70 0 0 1.000 5.5 C
2
TIME %A %B %C %D
FLOW PRESS
[
IN
]
0.0 30 70 0 0 1.000 5.5 C
2
TIME %A %B %C %D
FLOW PRESS
[
GR
]
0.0 40 60 0 0 1.000 5.5 C
2
TIME %A %B %C %D FLOW PRESS
[
LC
]
0.0
10 90
0
0 1.000
5.5 C 1
TIME %A %B %C %D
FLOW PRESS
[
IN
]
0.0
10 90
0
0 1.000
5.5
1
TIME %A %B %C %D
FLOW PRESS
[
FL
]
0.0 10 90 0 0 1.000 5.5 C
1