3 - 24
3.2.7 Column Flush Setting
The pump automatic stop function and the column flushing
function after the completion of analyses is set.
By setting of the column flush, the flushing will start
automatically at specified time after the analysis.
When a salt solution is used for analysis, the column must be
flushed with water to prevent elution of the salt component.
As an example, execution of the pump automatic stop function
or the column flushing function were set up after 15 minutes of
the analysis, and the column flushing time is set up as 30
minutes.
(to be continued)
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
3 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
AUTO PUMP OFF (YES=1, NO=0)
1
Select FLUSH
3 ENT
SET DELAY TIME (1.0 – 600.0) min
15.0
Select YES
1 ENT
SET UP
UTLTY
Monitor screen
Enter DELAY TIME for execution of the pump auto-stop
function and the column flush function from 1.0 to 600.0 min.
1 5 ENT
COLUMN FLUSH (YES=1, NO=0)
1
Select YES
1 ENT
FLUSH MODE (PARAM=1, PROGRAM NO.9=2)
1
1=flushing by the specified condition
2=flushing by the gradient program