The Kwik-Draw Pump is designed to measure
concentrations of gases and vapors when used with
AUER/MSA Detector Tubes.

Description

The Kwik-Draw Pump is a one-handed, manually-
operated bellows pump of 100 cc capacity.

Tube Holder

This rubber part permits mounting of detector tubes,
remote sampling lines or other detectors.

Filter Disc 

This porous plastic disc mounted in the rubber Tube
Holder protects the Pump from dirt and dust particles
which may alter the flow or damage the pump.

Exhaust Valve

Located under the valve cover, this valve closes as
the bellows re-inflates, and readily opens on the
exhaust stroke so blow-back through the tube holder
is negligible.

Stroke Counter

For convenience, a stroke counter is incorporated into
the Pump handle.

End-of-stroke Indicator

As the bellows begins to re-inflate, and after the knob
is released, the indicator eyeball turns high-visibility
green. As the vacuum decreases, the eye begins to
roll back to black. The stroke is over when the eye is
all black.

NOTE:  Kwik-Draw Pump (P/N 488543) does not have

an end-of-stroke indicator.

Operation

1.  Using the breaker on the Pump, break off both

tips of the Detector Tube.

2.  Using a twisting motion, insert the Tube into the

rubber tube holder. The arrow on the Tube

should point toward the Pump.

3.  Re-zero stroke counter.
4.  With all four fingers on the handle, depress the

knob with your palm.

NOTE: Watch the stroke counter; to ensure

proper sample volume, the counter 
will only advance if a full pump stroke 
is taken.

5. Release the knob.
6.  As the Pump re-inflates, the end-of stroke

indicator turns to high-visibility green. The 

stroke is over when the eye returns to the all

black state.

NOTE:  If your Pump does not have the end-of-

stroke indicator, wait 30 seconds after full
bellows inflation to ensure that all 100 cc
of the sample is drawn through the tube.
The Detector Tube must be held in the
sampling area during this period.

7.  To evaluate the stain, follow the instructions

provided with the Detector Tubes.

Remote Sampling

Remote sampling is accomplished by putting the
pump, connecting tube, remote sampling line and
Detector Tube together, in that order.

Maintenance

Under conditions of normal use, this Pump should
require little maintenance. Depending on the
frequency of use, periodic cleaning and checks for
correct performance are recommended. 

Tube Holder

Replace the Tube Holder when it shows signs of wear
or loss of elasticity. If filter is not clogged or cracked,
save the Filter Disc for re-use in the new Tube Holder.

Filter Disc

Periodically remove the Filter Disc for cleaning or
replacement.

1.  Remove the Filter Disc from the Tube Holder by

rolling the flange part of the Tube Holder down

and away from the Disc.

2.  Gently tap or blow on the surface to remove any

foreign matter.

3.  Replace Disc so previously exposed surface is

once again facing away from Pump.

Shaft

If the shaft is dirty or the bellows inflation is jerky,
remove the shaft by unscrewing; then, clean with 
auto wax.

Kwik-Draw

®

Sampling Pump 

Operation and Maintenance 

Deluxe Model with End-of-Stroke Indicator (P/N 487500)

Basic Model (P/N 488543) 