Test Results
398
88000 Series
Alarms/Errors Overview
RDI-L
(Remote Defect Indication - Line) - SONET
MS-RDI
(Multiplex Section - Remote Defect Indication) - SDH
RX
: Bits 6, 7, and 8 of the K2 byte contain the “110” pattern in five
consecutive frames.
TX
: Generates a “110” pattern for the bits 6, 7 and 8 of the K2 byte.
Errors
FAS-S
(Frame Alignment Signal - Section) - SONET
RS-FAS
(Regeneration Section - Frame Alignment Signal) - SDH
RX
: At least one A1 or A2 byte of the FAS word is in error.
B1
(BIP-8, Bit-Interleave Parity - 8 bits)
RX
: Indicates a Section (SONET) / Regeneration Section (SDH) parity
error by performing a routine even-parity check over all frames of the
previous STS-n/STM-n signal (located in the first STS-1/STM-1 of an
STS-n/STM-n signal).
B2
(BIP-8, Bit-Interleave Parity - 8 bits)
RX
:
SONET: Indicates a Line parity error by performing an even-parity
check over all bits of the LOH and SPE of the previous frame
(located in every STS-1 of an STS-n signal).
SDH: Indicates a Multiplex Section parity error by performing an
even-parity check over all bits (except those in the RSOH bytes) of
the previous frame of a STM-N signal.
REI-L
(Remote Error Indicator - Line) - SONET
MS-REI
(Multiplex Section - Remote Error Indicator) - SDH
RX
: The M0, M1, or the combination of both M0 and M1 bytes indicate
that one or more BIP violations have been detected. Refer to
on page 595 for more information. For OC-192, also
refer to
REI-L Computation Method
Summary of Contents for Power Blazer 88000 Series
Page 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Page 12: ......
Page 16: ......
Page 22: ......
Page 28: ......
Page 38: ......
Page 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Page 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Page 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Page 550: ......
Page 670: ......
Page 674: ......
Page 686: ......