Test Setup - Test Configurator, Timer, and System
144
88000 Series
BERT and Unframed BERT
Note:
Clearing the
Disruption Monitoring
check box will stop the measurement
without clearing the results. The disruption monitoring is automatically
stopped without clearing results when the test is stopped. However, starting
the test again while the
Disruption Monitoring
check box is selected will
reset the results before restarting.
Pass/Fail Verdict
check box when selected (cleared by default)
enables service disruption pass/fail verdict and allows setting the
threshold value.
SDT Threshold (ms)
allows entering the SDT threshold value that will
be used to declare the pass/fail verdict:
0.001
to
299999.999 ms
(default is
50
ms). For EtherBERT, the minimum value is adjusted with
respect to the
No Traffic Time
.
Multi-Pattern Configuration
Note:
O
nly available for
DSN/PDH BERT
test application with DS1 signal (DS0
disabled) and when the pattern is
Multi-Pattern
.
This feature allows sending each pattern for a specific duration,
sequentially and continuously.
Pattern
represents a sequence of patterns that will be generated:
1111
,
1in8
,
2in8
,
3in24
, and
QRSS
.
Enable
allows enabling the generation of each pattern individually in
the pattern sequence. All patterns are enabled by default. All 1’s (1111)
is disabled for
Unframed
test.
Individual Pattern Duration
specifies the transmission duration of
each pattern:
15 s
,
30 s
,
45 s
,
1 min
,
2 min
,
3 min
(default),... up to
15 min
.
Restore < Test Application > Defaults
Reverts the current test application to its default factory settings.
Summary of Contents for Power Blazer 88000 Series
Page 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Page 12: ......
Page 16: ......
Page 22: ......
Page 28: ......
Page 38: ......
Page 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Page 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Page 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Page 550: ......
Page 670: ......
Page 674: ......
Page 686: ......