Test Results
378
88000 Series
Alarms/Errors Overview
ODUx
Note:
For OPUx alarms see page 383. For GMP alarms/errors see page 373.
Alarms
AIS
(Alarm Indication Signal)
RX
: The STAT information in the PM byte 3, bits 6 to 8 is “111” for at
least 3 consecutive frames.
TX
: Generates an all "1"s pattern in the entire ODUk signal, excluding
the frame alignment overhead (FA OH), OTUk overhead (OTUk OH)
and ODUk FTFL.
BDI
(Backward Defect indication)
RX
: The BDI bit in the PM overhead field (byte 3, bit 5) is “1” for at least
5 consecutive frames.
TX
: Generates a “1” in the BDI (byte 3, bit 5) of the PM overhead field
continuously.
BSD
(Backward Signal Degrade)
RX
: The FTFL byte 128 is “00000010”.
TX
: Generates a “00000010” pattern in the FTFL Byte 128 continuously.
BSF
(Backward Signal Fail)
RX
: The FTFL byte 128 is “00000001”.
TX
: Generates a “00000001” pattern in the FTFL Byte 128 continuously.
FSD
(Forward Signal Degrade)
RX
: The FTFL byte 0 is “00000010”
TX
: Generates a “00000010” pattern in the FTFL Byte 0 continuously.
Summary of Contents for Power Blazer 88000 Series
Page 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Page 12: ......
Page 16: ......
Page 22: ......
Page 28: ......
Page 38: ......
Page 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Page 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Page 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Page 550: ......
Page 670: ......
Page 674: ......
Page 686: ......