Test Functions
588
88000 Series
OH - OTN
JC2
or
JC
1
JC2
, column 16, OPU OH: Controlled by GMP function. This byte is not
configurable.
JC
, column 16, OPU OH:
Bits 1-6, Reserved (RES), are configurable from binary
000000
(default) to
111111
.
Bits 7-8, Justification Control (JC), are configurable from binary
00
(default) to
11
. Not available with ODU mux. Changing the JC
value will corrupt the payload.
Row 3
TCM3/TCM2/TCM1
, columns 1-9, ODU OH: See
TCM6/TCM5/TCM4
on
page 587 for more information.
PM
, column 10-12, ODU OH: The Path Monitoring overhead contains
the following bytes.
The first PM byte (column 10) contains the TTI byte that is not
configurable.
The second PM byte (column 11) contains the BIP-8 byte and is
automatically generated for each frame. This byte is not configurable.
The third PM byte (column 12) contains the following sub-fields. This
byte is configurable from
00
to
FF
. The default value is
01
.
EXP
., column 13-14, ODU OH: The two Experimental overhead bytes
are configurable form
00
(default for each byte) to
FF
.
1. Depending on the test MUX/Mapping.
Bit 1-4
Bit 5
Bit 6-8
BEI
BDI
STAT
Summary of Contents for Power Blazer 88000 Series
Page 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Page 12: ......
Page 16: ......
Page 22: ......
Page 28: ......
Page 38: ......
Page 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Page 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Page 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Page 550: ......
Page 670: ......
Page 674: ......
Page 686: ......