Test Results
474
88000 Series
Summary (1588 PTP)
Summary (1588 PTP)
From the
Test
menu, tap
Results
, and the
Summary
tab.
Test Status
displays the current status of the 1588 PTP test. The global
test pass/fail verdict is displayed next to the
Test Status
field when
enabled (see
Start Time
indicates the date and time the test was started. The date
and time is reset every time the test is restarted or reset.
Test Recovery
, when displayed, indicates that the test has
automatically recovered from a power failure. The number of
occurrences is also displayed next to the
Test Recovery
field. Refer to
Logger Full
, when displayed, indicates that the logger is full. Refer to
Test Status
Description
--
Idle state, the test is not running or results are not available.
In Progress
The test is running.
Completed
The test is completed, stopped at the planned time, or manually
stopped when there is no set timer.
Aborted
The test is interrupted; stopped before the set timer.
For G.8265.1:
[Message type] Request Denied
,
Session Canceled
,
or
No Reply
is displayed next to
Aborted
when applicable.
For G.8275.1:
Sync Message Rate Changed
is displayed next to
Aborted
when applicable.
Summary of Contents for Power Blazer 88000 Series
Page 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Page 12: ......
Page 16: ......
Page 22: ......
Page 28: ......
Page 38: ......
Page 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Page 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Page 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Page 550: ......
Page 670: ......
Page 674: ......
Page 686: ......