Test Results
Power Blazer
379
Alarms/Errors Overview
FSF
(Forward Signal Fail)
RX
: The FTFL byte 0 is “00000001”.
TX
: Generates a “00000001” pattern in the FTFL Byte 0 continuously.
LCK
(Lock)
RX
: STAT information in the PM byte 3, bits 6 to 8 is “101” for at least 3
consecutive frames.
TX
: Generates a repeating "01010101" pattern in the entire ODUk signal,
excluding the frame alignment overhead (FA OH) and OTUk overhead
(OTUk OH).
LOFLOM
(Loss of Frame Loss Of Multiframe) is only available for
mapped ODU client signal (designated as LO in the standard).
RX
: OOF is present for at least 3 ms.
TX
: Generates error continuously in FAS and MFAS of a multiplexed
test case.
OCI
(Open Connection Indication)
RX
: STAT information in the PM byte 3, bits 6 to 8 is “110” for at least 3
consecutive frames.
TX
: Generates a repeating "01100110" pattern in the entire ODUk signal,
excluding the frame alignment overhead (FA OH) and OTUk overhead
(OTUk OH).
TIM
(Trace Identification Mismatch)
RX
: The received SAPI and/or DAPI do not math the expected SAPI
and/or DAPI. This alarm is only available when the SAPI ODU-TIM
and/or DAPI ODU-TIM check boxes are selected from
Summary of Contents for Power Blazer 88000 Series
Page 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Page 12: ......
Page 16: ......
Page 22: ......
Page 28: ......
Page 38: ......
Page 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Page 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Page 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Page 550: ......
Page 670: ......
Page 674: ......
Page 686: ......