Test Setup - Test Configurator, Timer, and System
Power Blazer
157
EtherBERT, FC BERT, BERT (CPRI/OBSAI), and Unframed BERT
BER Threshold
allows entering the threshold
Count
or
Rate
value that
will be used to declare the pass/fail verdict.
For
Count
, enter the maximum bit/pattern error count allowed before
declaring a fail verdict:
0
(default) to
999999
.
For
Rate
, enter the maximum bit/pattern error rate allowed before
declaring a fail verdict:
1.0E-14
to
1.9E-01
. The default value is
1.0E-12
;
2.1E-5
for
CLR4
unframed;
5.0E-5
for
SR4
,
CWDM4
, and
Other
unframed.
CPRI/OBSAI
Note:
Available with CPRI/OBSAI BERT test application with Framed L2.
Pass/Fail Verdict
check box when selected (default) enables the pass/fail
verdict. A pass verdict is given when the link status is up and no
CPRI/OBSAI alarm and/or error are recorded for the duration of the test.