Cal. No., point 4:
This can be set only if the calibration type is set to 4MB - 7MB or 4AB -
7AB.
Cal. No., point 5:
This can be set only if the calibration type is set to 5MB - 7MB or 5AB
- 7AB.
Cal. No., point 6:
This can be set only when the calibration type is set to 6MB - 7MB or 6AB
- 7AB.
Cal. No., point 7:
This can be set only if the calibration type is set to 7MB or 7AB.
Conc.:
Set the calibrator concentration value within 7 digits at each point.
Factor/OD-L:
This can be set only if the calibration type is set to ACAL (AB, AA, 2AB-7AB).
Set a factor range if the formula approximation is for a line. Set an absorbance range if the
formula approximation is for a curve.
Factor/OD-H:
This can be set only if the calibration type is set to ACAL (AB, AA, 2AB-7AB).
1-point cal. point:
This can be set only if the calibration type is set to 2AB - 7AB. Set a cal.
no. up to the maximum point that is specified according to the calibration type. This sets which
calibrator will be used for single point calibration.
W/Conc-O:
If a standard solution with a concentration value of 0 is being used, calibration can
be performed at a concentration value of 0 and at 2 other points. If a calibration uses 2 points
check this box.
Slope Check:
Slope Check has three settings:
“None” - The default setting.
“+” - Positive slope check compares all increasing OD values to the next calibration point.
“-” - Negative slope check compares all decreasing OD values to the next calibration point.
If an error occurs the user will hear an audible alarm and a error message will be generated.
Processing the Slope Check
When the AU2700 performs a Multi-point calibration, it compares the OD values for the setting
(“+” of “-”) selected. If the result of this comparison does not agree with the values selected, an
audible alarm is generated and the error message displayed below is displayed where “nn” is the
item number and “xxxx” is the test name that produced the error
4060 MULTI POINT-CAL. OD SLOPE CHECK ERROR (nn,xxxx)
If the slope check fails the result will print according to how the process is set in “Calibration
Specific Parameters” (OD or CONC).
MB type Factor:
This can be set only if the calibration type is set to MB. This is a specific
analyzer MB factor entered for the inner and outer wheel by your Beckman Coulter Technical
field service personnel.
Advanced Calibration:
Use this pull-down menu to program advanced calibration. For more
information on advanced calibration, refer to the Basic Operations Chapter.
Calibration Stability Period:
Enter the number of days, between 1 and 999. To leave this
field blank, enter a space. This is not needed for MB type assays.