C-56
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
????, 2006
2.8.2 Consumables
The following table lists consumable items and the consumption rate based on the number of
samples analyzed (the frequency of analysis).
Types of Consumables and the Consumption Rate
Name
Daily consumption
Wash solution B
(Used for washing cuvettes and mix bars)
Approx. 1.2 liters (8,000 tests/day)
(2% dilution)
ISE buffer solution
Approx. 180 mL (200 samples/day)
ISE MID solution
Approx. 260 mL (200 samples/day)
ISE REF solution
Approx. 35 mL (200 samples/day)
ISE cleaning solution
Approx. 1 mL/day (200 samples/day)
Tips
The ISE buffer, MID and REF solutions mentioned above are required for ISE operation. The solutions
are used during analysis, when performing periodic priming, and priming before analysis. The amount
of reagent used differs depending on the number of samples analyzed. The standards listed in the table
above are values calculated under the assumption that normal, routine work is being performed on the
analyzer.