Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-31
May, 2002
2.5.3 Entering Group (Batch) Requisitions (Routine, Emergency & Priority
STATS)
Perform a batch requisition when analyzing the same tests for multiple samples. For detailed
information on performing priority STAT requisition and analysis, refer to “Performing Priority
STAT Analysis.”
1. From the main menu, select [Routine], [Test Requisition] and [Normal], in that order.
The screen for requisitioning normal analysis is displayed.
OLP1017
Test Requisition Demographics
ALARMCLEAR
?
STANDBY
Help
Exit
S.No.
Delete
Sample
Normal
1997/09/03 10:10
ALB
GOT
GPT G-GTP
S.No.:
0001
Type:
Serum
Profile(P):
S.ID.:
Male
Sex:
Age:
(Year/Month)
Round No.:
Remaining Samples:
7100
Sample Dilution Rate:
Test
Total(T)
Search
ID(S)
1
XXXXXXXXXXXXXXXXXXXX
Start Entry
Change Test
Online
2. Select the following from the “Type” pull-down menu: serum, urine or other.
3. Verify the sample number is correct for the next sample number to be processed on the
AU2700.
If you are processing emergency samples: Type an “E” to test emergency
samples in the red rack or type a “P” to test emergency samples on the STAT table. Use
function key F8 (S. No.) to select a specific sample number.
4. Press function key F4 (start entry).
5. First, select tests individually or by a profile, and verify tests are highlighted. Individually:
Highlight the tests to be performed in blue by clicking them with your mouse. Tests
highlighted grey are not operational for the sample type selected. Profile: Click on the
profile button to open the “select profile window.” Select a profile number (1-99) and
click “OK.”
6. Press function key F3 (Xn-enter).
Enter the last sample number of the batch, or enter the
actual number of samples in the batch.
7. Click the “Execute” button.
The S. No. prompt advances to the next available sample
number for requisitioning. If the barcode reader is on, the sample ID numbers increment
by one.
8. Press function key F2 (exit).
The main screen returns.