Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-65
May, 2002
3.
Starting Analysis
This section describes the necessary procedures for starting analysis. Analysis should be started
from the Standby mode.
Starting Analysis
After preparing for analysis, perform the following procedure.
See Also
For information about preparing for analysis, refer to the Basic Operations Chapter.
1. On the main screen, select [Routine] and [Start Conditions] in that order.
The start
conditions are displayed.
OLP1019
ALARMCLEAR
?
STANDBY
Help
Exit
Start Condition
03/15/2000 10:10
Data Index:
Round No.:
Operator Name:
Serum Start No.
Normal
Repeat
03/12/2000
11 : 59
Routine
Routine Emergency STAT
Urine Start No.
Normal
Repeat
Others Start No.
Normal
Repeat
Routine Emergency STAT
Routine Emergency STAT
Calibration:
Remaining Samples: nnnn
Set
Cal
.
QC
Inventory
Check
Inventory
Check
Masking
2. Check the settings for the data index, round number, operator name. Also check the start
number for serum, urine, and other samples.
If the round number is changed, another
reagent check must be performed.
3. Verify that calibration status is “Selected” if performing reagent blanks or calibrations.
If
it is not selected, refer to “Performing Calibrations and a Reagent Blank” located in the
Preparing for Analysis section of this chapter.
4. Press function key F2 (exit).
The main screen is displayed.
5. Press the start key.
The window for starting analysis is displayed.