Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-71
May, 2002
Checking the Day-to-Day Variation Chart
1. On the main screen select [Routine], [QC Monitor] and [Day To Day Control] in that
order.
The screen for selecting the control sample for quality control is displayed.
OLP1037
ALARMCLEAR
?
STANDBY
Day to Day Control
1997/09/03 10:10
Help
Index:
Control Name:
yyyy / mm / dd hh : mm
yyyy / mm / dd hh : mm
-
serum
Type:
Exit
Day To Day
Change Test
Control No.
ITM01 ITM02 ITM03 ITM04 ITM05 ITM06 ITM07 ITM08 ITM09 ITM10 ITM11
(ALL)
Note
The index, sample type, and control name can be selected by using the pull-down menu located
by each box on the screen.
2. Press function key F8 (control no.). Enter the control number of the control sample to be
referenced and choose “OK.”
3. Select the test(s) to be referenced.
4. Press function key F6 (day to day).
The chart of day-to-day variation data is displayed.
OLP1038E
ALARMCLEAR
?
STANDBY
Day To Day Control
1997/09/03 10:10
Index:
Test Name:
Control Name:
xxxxxxxxxxxxxxxxxxxx
Mean
Day to Day Statistics:
N:
Mean:
SD:
CV(%):
Range:
Base Value:
Mean:
SD:
CV(%):
Range:
Help
Exit
Daily Data
Previous
Control/Test Control/Test
Next
Change Axis
nnn
nnnnnn.nnn
nnnnnn.nnn
nnnnnn.nnn
nnnnnn.nnn
nnnnnn.nnn
nnnnnn.nnn
nnnnnn.nnn
nnnnnn.nnn
xxxxxx
yyyy/mm/dd hh:mm
yyyy/mm/dd hh:mm
+3SD
+1SD
Ð1SD
Ð3SD
+2SD
Ð2SD
Type:
-
nnnnnnnn
nnnnnnnn
nnnnnnnn
nnnnnnnn
nnnnnnnn
nnnnnnnn
nnnnnnnn
3R
M
M
D
D
M
M
D
D
M
M
D
D
M
M
D
D
M
M
D
D
M
M
D
D
M
M
D
D
M
M
D
D
2R
1R
0
nnnnnnnn
nnnnnnnn
nnnnnnnn
nnnnnnnn
5. Check the chart for abnormal data.
See Also
For information about handling errors, refer to the Troubleshooting Chapter.