5.4 Periodic Maintenance
Updating the date of
execution by referring
to the state of periodic
maintenance
F6
Referring to the state
of periodic maintenance
execution for a data list
F7
F8
Registering periodic
maintenance items
Deleting periodic
maintenance items
F3
OLP3064
(M)Maintenance
(P)Periodic Maintenance
Outputting the
execution state of
periodic maintenance
in a data list.
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
STANDBY
Help
Exit
Execute
Periodic Maintenance
11/30/2000 10:10
OLP1pmain
No.
Maintenance Period Executed Next
1. Inspect Sample & Reagent Syringes
Daily 11/30/2000
12/01/2001
2. Inspect the Concentrated Wash Solution Level
Daily
5. Inspect/Clean the Sample Probes, Reagent Probes & Mix Bars
Daily
6. Inspect Printer & Paper
Daily
7. Prepare for a Sample Probe Wash
Daily
8. Change DI Water in the Pre-Dilution Bottle
Daily
9. W2
Weekly
10. Photocal
Weekly
11. Photometer Check
Weekly
12. Wash the Pre-dilution Bottle
Weekly
13. Clean All Wash Wells
Monthly
14. Clean the Wash Nozzle
Monthly
15. Clean the Deionized-Water Tank
Monthly
16. Clean the Deionized-Water Filter & Sample Probe Filter
Monthly
History
Entry
Delete
Grid
Display
Description:
Use this screen to determine which maintenance items are to be performed and if periodic
maintenance was executed on time. Use function key F4 (grid display) to display maintenance
in a table format.
No.:
Provides a number for the part listed.
Maintenance:
Displays the name of the maintenance item.
Period:
Displays the maintenance period in terms of days, weeks, months, and years.
Executed:
Displays the last date that the maintenance was performed.
Next:
Displays the date that maintenance is due.
Maintenance No.:
Displays the maintenance number to be edited. Use the </> button to
change the number.
Maintenance Name:
Type a name using 18 characters or less.
Period:
Set the maintenance period within the following range: Days - 1 to 1000, Weeks - 1 to
2500, Months - 1 to 600, Years - 1 to 50.
Period Unit:
Set the maintenance period in terms of days, weeks, months or years, using the
pull-down menu.