Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-47
May, 2002
2.2.5 Retrieve Repeat Data
F5
F6
OLP3035
Specifying the range
of repeat-run
samples to be
analyzed
Analyzing repeat-run
samples to register
them in a repeat-run
data file
(R) Routine
(T)Test Requisition
(S)Retrieve Repeat Data
Initializing a
repeat-run data file
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
WARM UP
Help
Exit
03/15/2000 10:10
OLP1rrd
Sample No.
Sample ID
Routine
Serum
Emergency
STAT
( nnnn - nnnn)
( Ennn - Ennn)
( Pnnn - Pnnn)
Routine
Urine
Emergency
STAT
( nnnn - nnnn)
(UEnnn - UEnnn)
Routine
Others
Emergency
STAT
( nnnn - nnnn)
(XEnnn - XEnnn)
(UPnnn - UPnnn)
(XPnnn - XPnnn)
Execute
Retrieve
File
Initialize
Retrieve Repeat Data
Description:
This screen has two main functions:
1. Initialize the repeat data if changing between auto repeat and manual repeat. This
initializes the repeat data file only.
2. This screen is also used to retrieve repeat requisitions with flags when using manual
repeat methods. Function key F5 (execute retrieve) is used when performing this
function.
Routine, Emergency, STAT (Serum, Urine, Others):
Select the sample type to be searched.
Each time the space bar is pressed, the check box switches from “Checked” to “Unchecked.”
Sample No.:
Enter the sample number range in this field. Enter an asterisk (*) to designate all
samples.
Sample ID:
Enter the sample ID range according to the range set in [Parameters], [Requisition
Format]. Enter an asterisk (*) to designate all sample IDs