D-28
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
1.9 Data Processor (DPR) Status
Referring to the data
for samples under
analysis
F5
Referring to main
information on
system status
[System Status]
(4) Data
Display
Referring to the data
for all samples
Referring to the data
for ISE (option)
Referring to the data
for samples analyzed
with the STAT table
Select [All Data] tab
"Data Display" screen
Select [STAT Data] tab
Select [ISE Data] tab
F6
F5
Referring to data
related to analyzed
smples
F5
OLU3013E
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
WARM UP
Help
Exit
03/15/2000 10:10
OLP1stdpr
Normal Run
Repeat Run
Serum
Urine
Others
Serum
Urine
Others
Serum
Urine
Others
Routine Requisition Data Requisition Data
Emergency
STAT
Total Samples:
0 0 0 0
System Status - DPR Status
Reagent Blank:
Calibration:
QC:
Data
Measure Status
Online Status
Input
S. No.:
Output
S. No.:
Printer Status
S. No.:
System
Status Icon
Printer
Control
Description:
This screen displays the sample number for which analysis has started and ended. It also
displays sample numbers that were input/output on-line between the system and host computer,
and the status of the printer.
Use this software screen to:
View sample analysis, data transfer, and printer status:
1. Select [System Status], [DPR Status].
Measure Status:
Normal run and repeat run
Displays the sample number for which analysis has started (requisition column) and ended (data
column). If there are no corresponding samples, nothing is displayed.
Total number of samples
The total number of samples for which analysis has been started and completed displays. If no
samples were run, “0” displays.
Online Input/Output Status:
Input/output
The selected method of input or output (ONLINE) and “Real-time” or “Batch” will be displayed.
Sample number
Sample numbers for which the data was entered or printed displays.
Printer Status
:
Real-time, report, or batch printing
While in real-time printing, “Real-time printing” displays. While in report printing, “Report
printing” displays. In other printing states, “Batch printing” displays.
Sample number
The sample number printed displays while in real-time printing or report printing.
Содержание AU2700
Страница 17: ...A 4b Automated Chemistry Analyzer AU2700 User s Guide November 1 2009 This page intentionally left blank ...
Страница 178: ...D 4 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide March 31 2005 ...
Страница 260: ...D 86 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide May 2002 ...