Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-153
February 4, 2004
Edit a log-in name:
To edit a log-in name or menu level, select function key F5 (edit). Make your changes then
select "Close."
Delete a new log-in name:
To delete a log-in name, select function key F6 (delete).
A window will appear confirming the
deletion.
Select "OK."
Tip
When Login Function is enabled, software menus default to level one security. At least one Login name
must have level one security assigned or menus and sub menus will be inaccessible.
Change security level of individual menus and submenus:
1. Select [Parameters], [System], [Login Name].
2. Select the 'Menu Level' tab.
3. Select F5: Edit.
4. Enter the menu access level for each menu.
The default menu level is 1, which is the
highest security level.
Be sure to program menus required for basic operation at a level
10 so all operators have access.
5. Select F2: Exit. The message "This function will not be enabled until next Power On"
displays.
6. Perform an End Process.
7. Press the ON button.
8. The Login Window appears after F2: Exit is selected from [Start Condition].