D-120
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
D-120
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
March 9, 2007
Type:
The sample type can be switched. Select either “Serum”, “Urine”, or “Other” from the
pull-down list.
Use serum cal.:
If this box is selected, the data is calculated in accordance with serum
calibration during sample analysis. This box displays only if urine sample type is selected.
Calibration type:
Select the calibration type for each analysis test from the pull-down menu.
Formula:
Set a formula according to the calibration type using the pull-down menu.
OD:
Set the calibrator OD value within the range of -2.0 to 2.5 at each point.
Counts:
Set the number of times to repeatedly perform calibration analysis and a reagent blank
analysis. Enter a number between 1 and 4. Depending on the number, the following values will
be used:
Set to 1: The OD value is used.
Set to 2: An average of two values.
Set to 3: An average of two close values.
Set to 4: An average of two values, excluding maximum and minimum values.
Process:
Set the method to be used if an update of the calibration fails. Set to either “OD”
or “CONC.” To perform a concentration conversion using the existing factors, set to
“CONC.” Otherwise, set to “OD” if a concentration conversion is not initially performed.
Save the measured OD value, and batch-perform the concentration conversion after
performing the calibration again. This can be set only if the calibration type is set to ACAL
(AB, AA, 2AB-7AB).
Cal.No. at each Calibrator specification point:
Enter the calibration number from 1 to 200
defined in the [Parameters], [Calibration Parameters], [Calibrator] menu.
Tip
A maximum of two digits can be entered to specify a calibrator number. If specifying a range between 1
and 99, enter the two numeric values as they are. If specifying a range between 100 and 200, enter the
alphanumeric value corresponding to the chart on page D-118 of the User Guide.
Cal. No., point 1:
This can be set only if the calibration type is set to 2MB - 7MB, AB, AA or
2AB-7MB.
Cal. No., point 2
: This can be set only if the calibration type is set to 2MB - 7MB or 2AB -
7AB or AA.
Cal. No., point 3:
This can be set only if the calibration type is set to 3MB - 7MB or 3AB -
7AB.